Limstar - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

LIMSTAR

LIMSTAR
Syntax
LIMSTAR <numeric>[HZ|MHZ|V|A]
LIMSTAR?
Description
In the creation/edit of the limit line table, sets the sweep parameter value of each segment
start point.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the sweep parameter value of the start
point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the sweep parameter value of the start point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[System]
key
334
Description
Sweep parameter value of the segment start point
Range
same as "STAR" command on page 444
Initial value
When creating segment 1: same as "STAR" command
When creating an additional segment: Set value of the previous segment
Unit
same as "STAR" command
Resolution
same as "STAR" command
-
LIMIT TEST
-
<numeric>
EDIT LIMIT LINE
-
EDIT - START
Chapter 16