GPIB Command Reference

SEANPKR

SEANPKR
Syntax
SEANPKR
Description
For the active trace (set with the "TRAC" command on page 457), moves the marker to a
peak to the right of the current marker position. If no peak is detected, the message saying
"Not detected" is displayed (no error occurs). If the marker function is off, executing this
command causes an error and the command is ignored. (No query)
Corresponding
[Search]
key

SEAR

Syntax
SEAR
Description
For the active trace (set with the "TRAC" command on page 457), moves the marker to the
target (set with the "SEATARG" command on page 433) to the right of the current marker
position. If no target is detected, the message saying "Not detected" is displayed (no error
occurs). If the marker function is off, executing this command causes an error and the
command is ignored. (No query)
Corresponding
[Search]
key

SEARLINE

Syntax
SEARLINE {ON|OFF|1|0}
SEARLINE?
Description
For the active trace (set with the "TRAC" command on page 457), enables/disables the
display of the lines indicating the search range at list sweep.
Parameters
Query response
{1|0}<newline><^END>
Corresponding
[Search]
key
428
-
-
PEAK
NEXT PEAK RIGHT
-
-
TARGET
SEARCH LIGHT
ON or 1
Enables the display of the lines indicating the search range.
OFF or 0
Disables the display of the lines indicating the search range.
(initial value)
-
SEARCH RANGE MENU
Description
-
RNG LINE on OFF