Reading/Writing Measurement Data
Reading/Writing Data
Line 870 to 890
Specifies trace B as the active trace and sets the maximum and
minimum value of the screen for trace B to Top_b and Btm_b,
respectively.
Lines 900 to 1030 Creates a limit line table for trace B and turns ON the limit test
function.
Lines 1050 to 1070 Performs a sweep once and waits for the sweep to be completed.
Lines 1090 to 1110 Specifies trace A as the active trace and then reads the number of
measurement points for trace A where the measurement data were
found to be unacceptable.
Lines 1130 to 1230 Specifies the ASCII format as the data transfer format and reads and
displays the sweep parameter value, test result, and upper and lower
limits for each of the measurement points for trace A where the
measurement data were found to be unacceptable.
Line 1250
Displays a message if the measurement data at all measurement points
for trace A were found to be acceptable.
Lines 1280 to 1300 Specifies trace B as the active trace and then reads the number of
measurement points for trace B where the measurement data were
found to be unacceptable.
Lines 1320 to 1420 Specifies the ASCII format as data transfer format and reads and
displays the sweep parameter value, test result, and upper and lower
limits for each of the measurements points for trace B where the
measurement data were found to be unacceptable.
Line 1440
Displays a message if the measurement data at all measurement points
for trace B were found to be acceptable.