Example 6-4
Reading Limit Test Results
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
270
280
290
300
310
320
330
340
350
360
370
380
390
400
410
420
430
440
450
460
470
480
490
500
510
520
530
540
550
560
570
580
590
600
610
Chapter 6
REAL Star,Stop,Top_a,Btm_a,Top_b,Btm_b
REAL Fail_res_a(1:41,1:4),Fail_res_b(1:41,1:4)
REAL Lim_star_a(1:3),U_lim_star_a(1:3),L_lim_star_a(1:3)
REAL Lim_stop_a(1:3),U_lim_stop_a(1:3),L_lim_stop_a(1:3)
REAL Lim_star_b(1:3),U_lim_star_b(1:3),L_lim_star_b(1:3)
REAL Lim_stop_b(1:3),U_lim_stop_b(1:3),L_lim_stop_b(1:3)
DIM Meas_para$[9],Buff$[9]
INTEGER Nop,Fail_nop
CLEAR SCREEN
IF SYSTEM$("SYSTEM ID")="HP4294A" THEN
ASSIGN @Hp4294a TO 800
ELSE
ASSIGN @Hp4294a TO 717
END IF
Meas_para$="CSD"
Star=1.0E+7
Stop=5.0E+7
Nop=41
Top_a=2.0E-5
Btm_a=0.
Top_b=.02
Btm_b=0.
!
Lim_star_a(1)=1.0E+7
U_lim_star_a(1)=1.2E-5
L_lim_star_a(1)=8.E-6
Lim_stop_a(1)=2.0E+7
U_lim_stop_a(1)=1.2E-5
L_lim_stop_a(1)=8.E-6
Lim_star_a(2)=2.0E+7
U_lim_star_a(2)=1.2E-5
L_lim_star_a(2)=8.E-6
Lim_stop_a(2)=4.0E+7
U_lim_stop_a(2)=1.5E-5
L_lim_stop_a(2)=5.E-6
Lim_star_a(3)=4.0E+7
U_lim_star_a(3)=1.5E-5
L_lim_star_a(3)=5.E-6
Lim_stop_a(3)=5.0E+7
U_lim_stop_a(3)=1.5E-5
L_lim_stop_a(3)=5.E-6
!
Lim_star_b(1)=1.0E+7
U_lim_star_b(1)=.005
L_lim_star_b(1)=0.
Lim_stop_b(1)=2.0E+7
U_lim_stop_b(1)=.005
L_lim_stop_b(1)=0.
Lim_star_b(2)=2.0E+7
U_lim_star_b(2)=.005
L_lim_star_b(2)=0.
Lim_stop_b(2)=4.0E+7
U_lim_stop_b(2)=.01
L_lim_stop_b(2)=0.
Lim_star_b(3)=4.0E+7
U_lim_star_b(3)=.01
L_lim_star_b(3)=0.
Lim_stop_b(3)=5.0E+7
U_lim_stop_b(3)=.01
L_lim_stop_b(3)=0.
!