Dcomopen{G|C - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

DCOMOPEN{G|C}

Syntax
DCOMOPENG <numeric>[SIE]
DCOMOPENC <numeric>
DCOMOPEN{G|C}?
Description
For fixture compensation data measurement ("COM{A|B|C}" command on page 281), sets
the conductance value ("DCOMOPENG" command) and the capacitance value
("DCOMOPENC" command) of the OPEN standard used to measure OPEN compensation
data.
Parameters
Description
Range
Initial value
Unit
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Cal]
key
Chapter 16
Conductance value or capacitance value
*1
-1E6 to 1E6
0
For conductance value: S (siemens)
For capacitance value: fF (femto farad)
For conductance value: 1E-12
For capacitance value: 1
*1.The unit of range value is S (siemens) or F (farad).
-
FIXTURE COMPEN
-
DEFINE VALUE
GPIB Command Reference
DCOMOPEN{G|C}
<numeric>
- {
OPEN CONDUCT (G)
|
OPEN CAP (C)
}
293