OUTPDTRCP?

Syntax
OUTPDTRCP? <numeric 1>
Description
Reads out the value of the specified measurement point in a data trace array (refer to
"Internal data arrays" on page 81). (Query only)
Parameters
Description
Range
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 2},{numeric 3}<newline><^END>
Reads out the readout and subsidiary readout of the measurement parameter value as
shown below.
{numeric 2}:
{numeric 3}:
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
1 to the number of points
1
Readout of the specified measurement point. If the measurement
parameter is a scalar value (for other than COMPLEX Z-Y), the
measurement parameter value is read out. If the measurement
parameter is a vector value (for COMPLEX Z-Y), resistance (for trace
A) or conductance (for trace B) is read out.
Subsidiary readout of the specified measurement point. If the
measurement parameter is a scalar value (for other than COMPLEX
Z-Y), 0 is always read out. If the measurement parameter is a vector
value (for COMPLEX Z-Y), reactance (for trace A) or susceptance
(for trace B) is read out.
GPIB Command Reference
OUTPDTRCP?
<numeric 1>
379