GPIB Command Reference

LIMDSTOP

LIMDSTOP
Syntax
LIMDSTOP <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
LIMDSTOP?
Description
In the creation/edit of the limit line table, sets the limit width (=upper value - central value)
of the stop point of each segment (specified with the "LIMSTOP" command on page 336).
Use this command, together with the "LIMVSTOP" command on page 340, when
specifying the limits using the central value and the width.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the limit width of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the limit width of the stop point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[System]
key

LIMEDONE

Syntax
LIMEDONE
Description
Finishes the creation/edit of the limit line table. Executing this command brings the
created/edited table to take effect. (No query)
Corresponding
[System]
key
326
Description
Limit width of the segment stop point
Range
-1E9 to 1E9
Initial value
When creating segment 1: 0
When creating an additional segment: Set value of the previous segment
Unit
Varies depending on the measurement parameter (refer to the explanation
of unit in "SCAL" on page 423).
-
-
LIMIT TEST
-
-
LIMIT TEST
<numeric>
-
EDIT LIMIT LINE
EDIT
-
EDIT LIMIT LINE
done
-
-
more 1/3
DELTA LIMIT