Dcomshor{R|L; Dcrng - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

DCOMSHOR{R|L}

DCOMSHOR{R|L}
Syntax
DCOMSHORR <numeric>[OHM]
DCOMSHORL <numeric>[H]
DCOMSHOR{R|L}?
Description
For fixture compensation data measurement ("COM{A|B|C}" command on page 281), sets
the resistance value ("DCOMSHORR" command) and the inductance value
("DCOMSHORL" command) of the SHORT standard used to measure SHORT
compensation data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Cal]
key

DCRNG

Syntax
DCRNG {M1|M10|M100}
DCRNG?
Description
Sets the dc bias range.
Parameters
Query response
{M1|M10|M100}<newline><^END>
Corresponding
[Source]
key
294
Description
Resistance value or inductance value
Range
-10E6 to 10E6
Initial value
0
For resistance value: Ω (ohm)
Unit
For inductance value: H (henry)
Resolution
1E-12
-
FIXTURE COMPEN
M1 (initial value)
Specifies the 1-mA range.
M10
Specifies the 10-mA range.
M100
Specifies the 100-mA range.
-
-
more 1/2
MEAS RANGE [ ]
<numeric>
-
- {
DEFINE VALUE
SHORT RESIST(R)
Description
- {
|
1mA
10mA
|
SHORT INDUCT(L)
|
}
100mA
Chapter 16
}