GPIB Command Reference

LMINS?

LMINS?
Syntax
LMINS? <numeric 1>
Description
Reads out the measurement parameter value and sweep parameter value of the negative
peak at the location specified with the order from the left edge of the sweep range among
all negative peaks within the waveform analysis range set with the "ANARANG"
command on page 267.
The negative peak searched for at the execution of this command satisfies the condition set
with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Parameters
If the specified parameter is out of the setting range, the operation is the same as when no
peak is detected.
Query response
{numeric 2},{numeric 3}<newline><^END>
If the specified negative peak is not detected, the readouts of {numeric 2} and {numeric 3}
are 3.40282346639E+38 and 0, respectively.
Corresponding
No front panel key is available to execute this function.
key
344
Description
Location of the negative peak (the order of the negative peak from
the left)
Range
1 to the number of points
Resolution
1
Description
Measurement parameter value
<numeric 1>
{numeric 2}
Sweep parameter value
{numeric 3}