EQUC

Syntax
EQUC {CIRA|CIRB|CIRC|CIRD|CIRE}
EQUC?
Description
Selects an equivalent circuit model (A to E) used when executing the equivalent circuit
analysis with the "CALECPARA" command on page 275. For details on each equivalent
circuit model, refer to "Equivalent circuit analysis" on page 100.
Parameters
CIRA (initial value) Specifies equivalent circuit model A.
CIRB
CIRC
CIRD
CIRE
Query response
{CIRA|CIRB|CIRC|CIRD|CIRE}<newline><^END>
Corresponding
[Display]
key

EQUC0?

Syntax
EQUC0? <numeric 1>[HZ|MHZ]
Description
If the sweep parameter is frequency, within the waveform analysis range specified with the
"ANARANG" command on page 267, reads out equivalent circuit parameter C0 (parallel
capacitance) at the specified frequency. If the sweep parameter is not frequency, executing
this command causes an error and 0 is always read out.(Query only)
Parameters
Description
Range
Unit
Resolution
Query response
{numeric 2}<newline><^END>
Description
Unit
Note that, if the specified frequency is out of the analysis range, 0 is read out.
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
Specifies equivalent circuit model B.
Specifies equivalent circuit model C.
Specifies equivalent circuit model D.
Specifies equivalent circuit model E.
-
more 1/2
-
EQUIV CKT
-
SELECT CIRCUIT
Frequency you want to specify
40 to 110E6
Hz
1E-3
C0 (parallel capacitance) at the specified frequency
F (farad)
GPIB Command Reference
Description
- {
A
|
B
|
C
|
D
|
E
}
<numeric 1>
{numeric 2}