LIMSADD

LIMSADD
Syntax
LIMSADD [<numeric>]
Description
In the creation/edit of the limit line table, adds a new segment, changes the edit target
segment to the added segment, and starts edit. The position where a segment is added is as
follows:
The number of the added segment is a value of the specified segment number (if no
segment is specified, the edit target segment number before the execution of the command)
plus 1, and the number of each following segment also increases by 1.
After the completion of the parameter setting of the segment, execute the "LIMSDON"
command on page 331 to finish the edit work. After executing this command and the
"LIMSEDI" command on page 332 (to start segment edit), executing this command,
without executing the "LIMSDON" command, causes an error and the command is
ignored.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit and then adds a segment. (No query)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Corresponding
[System]
key
330
If a segment is specified:
Added in the line below the specified segment.
If no segment is specified:
Added in the line below the segment specified as the edit target (set with the
"LIMSEGM" command on page 333).
Description
Segment number immediately before the addition and insertion
location
Range
1 to the number of the segments in the limit line table
Resolution
1
-
LIMIT TEST
-
<numeric>
ADD 