Limsedi - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

LIMSEDI

LIMSEDI
Syntax
LIMSEDI <numeric>
Description
In the creation/edit of the limit line table, starts the edit of the segment specified as the edit
target (set with the "LIMSEGM" command on page 333). Note that, if a segment is
specified, the edit target segment is changed to the specified segment and then the edit is
started.
After the completion of the parameter setting of the segment, execute the "LIMSDON"
command on page 331 to finish the edit work. After executing this command and the
"LIMSADD" command on page 330 (to start segment edit), executing this command,
without executing the "LIMSDON" command, causes an error and the command is
ignored.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit and then starts the edit of the segment. (No query)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Corresponding
[System]
key
332
Description
Segment number you want to edit
Range
1 to the number of the segments in the limit line table
Resolution
1
-