Negl; Nexnpk - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

NEGL

NEGL
Syntax
NEGL
NEGL?
Description
Sets the input/output signals of the 24-bit I/O port (data and test PASS/FAIL output) to
negative logic (preset state). To specify positive logic, use the "POSL" command on page
403.
Query response
{1|0}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key

NEXNPK?

Syntax
NEXNPK?
Description
Within the waveform analysis range set with the "ANARANG" command on page 267,
searches for the minimum peak having a value more than the value of the negative peak
that has been detected at the last execution of this command or the "NPEAK?" command
on page 364, and reads out the measurement parameter value and sweep parameter value of
the negative peak. If there are several applicable negative peaks, reads out the value of the
leftmost negative peak within the analysis range. If no peak exists, 0 is read out.
The negative peak searched for at the execution of this command satisfies the condition set
with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response
{numeric 1},{numeric 2}<newline><^END>
Readouts are as follows:
{numeric 1}:
{numeric 2}:
Corresponding
No front panel key is available to execute this function.
key
362
1
Negative logic is specified.
0
Negative logic is not specified. In other words, positive logic is specified.
Measurement parameter value of the detected negative peak
Sweep parameter value of the detected negative peak
Description