Agilent Technologies 4294A Programming Manual page 209

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

2450
!
2460
OUTPUT @Agt4294a;";STAR ";Start
2470
OUTPUT @Agt4294a;";STOP ";Stop
2480
OUTPUT @Agt4294a;";SWPT ";Swe_type$
2490
OUTPUT @Agt4294a;"POIN ";Nop
2500
OUTPUT @Agt4294a;";BWFACT ";Bw
2510
OUTPUT @Agt4294a;";POWMOD VOLT;POWE ";Vosc
2520
OUTPUT @Agt4294a;"DCOMOPENG ";Load_g
2530
OUTPUT @Agt4294a;"DCOMOPENC ";Load_c/1.E-15
2540
OUTPUT @Agt4294a;"DCOMSHORR ";Short_r
2550
OUTPUT @Agt4294a;"DCOMSHORL ";Short_l
2560
OUTPUT @Agt4294a;"DCOMLOADR ";Open_r
2570
OUTPUT @Agt4294a;"DCOMLOADL ";Open_l
2580
!
2590
Result=FNUser_Corr(@Agt4294a,"Compen_Open")
2600
IF Result<>0 THEN Prog_end
2610
!
2620
!
2630
DISP "Set 16451B OPEN condition and press 'Continue'."
2640
PAUSE
2650
Result=FNUser_Corr(@Agt4294a,"Compen_Load")
2660
IF Result<>0 THEN Prog_end
2670
!
2680
!
2690
DISP "Set 16451B SHORT condition and press 'Continue'."
2700
PAUSE
2710
Result=FNUser_Corr(@Agt4294a,"Compen_Short")
2720
IF Result<>0 THEN Prog_end
2730
!
2740
!
2750
DISP ""
2760
PRINT TABXY(2,7);"x]
2770
RETURN
2780
!
2790 Input_t:!
2800
DISP "Thickness of MUT [mm] = ?";
2810
INPUT "",Ans
2820
T_dut=Ans/1000
2830
RETURN
2840
!
2850 Input_tg:!
2860
DISP "Distance of Electrode Gap Tg [mm] = ?";
2870
INPUT "",Ans
2880
Tg=Ans/1000
2890
RETURN
2900
!
2910 Input_d:!
2920
SELECT Elec$
2930
CASE "A"
2940
D_elec=D_elec_a
2950
CASE "B"
2960
D_elec=D_elec_b
2970
END SELECT
2980
RETURN
2990
!
3000 Sweep: !
3010
OUTPUT @Agt4294a;"DISA BASS"
3020
OUTPUT @Agt4294a;"MEAS CPD"
3030
OUTPUT @Agt4294a;"HIDI OFF"
3040
OUTPUT @Agt4294a;"SING"
3050
DISP "Now measuring Cp-D..."
3060
OUTPUT @Agt4294a;"*OPC?"
3070
ENTER @Agt4294a;Opc
3080
DISP "Now getting data..."
13
Application Sample Programs
Measuring Dielectric Material
"
209

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents