GPIB Command Reference
DCALOPEN{G|C}
DCALOPEN{G|C}
Syntax
DCALOPENG <numeric>[SIE]
DCALOPENC <numeric>
DCALOPEN{G|C}?
Description
For user calibration data measurement ("CAL{A|B|C}" command on page 274), sets the
conductance value ("DCALOPENG" command) and the capacitance value
("DCALOPENC" command) of the OPEN standard of the calibration kit used to measure
OPEN calibration data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[CAL]
key
288
Description
Conductance value or capacitance value
*1
-1E6 to 1E6
Range
Initial value
0
Unit
For conductance value: S (siemens)
For capacitance value: fF (femto farad)
Resolution
For conductance value: 1E-12
For capacitance value: 1
*1.The unit of range value is S (siemens) or F (farad).
-
-
USER CAL
DEFINE VAUE
<numeric>
- {
OPEN CONDUCT (G)
|
}
OPEN CAP (C)
Chapter 16