GPIB Command Reference

OUTPMWID?

OUTPMWID?
Syntax
OUTPMWID?
Description
For the active trace (set with the "TRAC" command on page 457), if the trace bandwidth
analysis function is on (specified to ON with the "WIDT" command on page 461), reads
out the result of the bandwidth search. (Query only)
Query response
{numeric 1},{numeric 2},{numeric 3},{numeric 4},{numeric 5},{numeric 6}<newline><^END>
The readouts are as follows:
{numeric 1}:
{numeric 2}:
{numeric 3}:
{numeric 4}:
{numeric 5}:
{numeric 6}:
If the trace bandwidth analysis function is off, the readouts are invalid.
Corresponding
No front panel key is available to execute this function.
key
390
Bandwidth.
Center frequency.
Quality factor.
Measurement parameter value of the maker position.
Value of the left cutoff point sweep parameter value minus the center
value of the sweep range.
Value of the right cutoff point sweep parameter value minus the center
value of the sweep range.
Chapter 16