OSPT
Syntax
OSPT <numeric>
OSPT?
Description
Sets the value of the Positive Transition Filter of the Operation Status Register.
Parameters
Description
Range
Initial value
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key
OSR?
Syntax
OSR?
Description
Reads out the value of the Operation Status Condition Register. (Query only)
Corresponding
No front panel key is available to execute this function.
key
OUT1{H|L}
Syntax
OUT1{H|L}
Description
Sets OUTPUT1 of the 24-bit I/O port to HIGH (or LOW). (No query)
Corresponding
No front panel key is available to execute this function.
key
OUT1ENV{H|L}
Syntax
OUT1ENV{H|L}
OUT1ENV{H|L}?
Description
Makes the setting so that OUTPUT1 goes to HIGH (or LOW) when a pulse is inputted to
INPUT1 of the 24-bit I/O port.
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
<numeric>
Value of the filter
0 to 65535
0
1
GPIB Command Reference
OSPT
367