OUTPMEMO?

Syntax
OUTPMEMO?
Description
Reads out the values (complex number) of all measurement points in the memory array
(refer to "Internal data arrays" on page 81). (Query only)
Query response
{numeric 1},{numeric 2},..,{numeric NOP×2-1},{numeric NOP×2}<newline><^END>
Reads out the real part and imaginary part of each measurement point value (complex
number) separately. Where, NOP is the number of points, and n is an integer between 1 and
NOP.
{numeric n×2-1}: Measurement parameter value of the n-th measurement point (real
{numeric n×2}:
Corresponding
No front panel key is available to execute this function.
key

OUTPMEMOP?

Syntax
OUTPMEMOP? <numeric 1>
Description
Reads out the value (complex number) of the specified measurement point in the memory
array (refer to "Internal data arrays" on page 81). (Query only)
Parameters
Description
Range
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 2},{numeric 3}<newline><^END>
The real part and imaginary part are read out separately as shown below.
{numeric 2}:
{numeric 3}:
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
part).
Measurement parameter value of the n-th measurement point
(imaginary part).
Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
1 to the number of points
1
Measurement parameter value of the specified measurement point
(real part).
Measurement parameter value of the specified measurement point
(imaginary part).
GPIB Command Reference
OUTPMEMO?
<numeric 1>
385