Inpucomc{1-3; Inpudata - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

INPUCOMC{1-3}

Syntax
INPUCOMC{1-3} <numeric 1>,<numeric 2>,..,<numeric NOP×2-1>,<numeric NOP×2>
Description
Enters the OPEN/SHORT/LOAD compensation data for fixture compensation, calculates
the compensation coefficient, and stores it into the non-volatile memory (backup memory).
The "INPUCOMC1" command is for OPEN compensation data (G-B); the
"INPUCOMC2" command, SHORT compensation data (R-X); the "INPUCOMC3"
command, for LOAD compensation data (R-X).
Enter data after setting the standard values, because the standard values to be used in
measuring data is needed to calculate fixture compensation coefficients.
A mismatch, between the number of required data at the execution of the command
(defined by the measurement point setting for fixture compensation data (set with the
"CALP" command on page 275) and the number-of-points setting) and the number of
entered parameters, causes an error and the command is ignored. (No query)
Parameters
<numeric n×2-1>
<numeric n×2>
Total of NOP×2 parameters are required because the fixture compensation data value
(complex number) of each measurement point is divided into 2 parameters for its real part
and imaginary part. Where, NOP is the number of points (for user measurement points) or
100 (for fixed measurement points), and n is an integer between 1 and NOP.
Corresponding
No front panel key is available to execute this function.
key

INPUDATA

Syntax
INPUDATA <numeric 1>,<numeric 2>,..,<numeric NOP×2-1>,<numeric NOP×2>
Description
Enters a data array (refer to "Internal data arrays" on page 81). A mismatch, between the
number-of-points setting at the execution of the command and the number of entered
parameters, causes an error and the command is ignored. (No query)
Parameters
<numeric n×2-1>
<numeric n×2>
Total of NOP×2 parameters are required because the measurement parameter value of each
measurement point (complex number) is divided into 2 parameters for its real part and
imaginary part. Where, NOP is the number of points, and n is an integer between 1 and
NOP.
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
Description
Real part of the compensation data value (complex number) of the
n-th measurement point.
Imaginary part of the compensation data value (complex number) of
the n-th measurement point.
Description
Real part of the measurement parameter value (complex number) of
the n-th measurement point.
Imaginary part of the measurement parameter value (complex
number) of the n-th measurement point.