"Laser Active" Signals

The signals for "laser active" operation – after
set_laser_control
– can be activated via
set_laser_pulses,
set_laser_pulses_ctrl
set_laser_timing. Additionally they can be enabled
via
set_laser_control
or
enable_laser
via
set_laser_control
or
disable_laser
command
get_startstop_info
current status of the laser control signals (enabled or
disabled).
Even if the laser control signals for "laser active"
operation have been enabled and activated as
described above, they will only appear at the output
ports if they are switched on via further commands.
They are (automatically) switched on when mark or
arc commands are executed (see
also be switched on for a fixed scanner position via
the command
laser_on_list
(2)
interval
or switched on (for an indefinite period of
time) via
laser_signal_on
or
and switched off via
laser_signal_off
laser_signal_off_list.
The laser control signals for "laser active" operation
are automatically switched off by the RTC
marking command (mark or arc) is followed by a non-
marking command, when a list is stopped (via
set_end_of_list
or stop_execution) or a list is
temporarily interrupted (after set_wait,
or stop_list). In the latter case, the signals are
switched on again, when the list is continued (after
release_wait
or restart_list). The laser control
signals are also automatically switched off by a
list_nop
command. This command will be automati-
®
cally inserted by the RTC
5 under some circum-
stances, e.g. as a replacement for list commands with
an invalid parameter.
(1) If the signals are disabled, the laser output ports (LASERON,
LASER1 and LASER2) are in the (high impedance) tristate mode
(the signals will be automatically disabled by pause_list, too,
and reenabled by restart_list).
(2) Via laser_on_pulses_list, you can also switch on the LASERON
signal for a specified number of external signal pulses.
®
RTC
5 PC Interface Board
Rev. 1.9 e
7 Basic Functions for Scan Head and Laser Control
or
and disabled
(1)
. The
(Bit #9) queries the
page
93); they can
for a specified time
laser_signal_on_list
or
®
5, when a
pause_list

"Laser Standby" Signals

After first-time activation of the laser control signals
via set_laser_control, standby signals can be acti-
vated and deactivated via
set_standby
set_standby_list.
Standby signals will be permanently outputted after
activation without further commands if and only if no
signals for "laser active" operation are switched on.
If applicable, the commands which switch on and
switch off the signals for "laser active" operation (see
previous section), correspondingly switch from
standby to "laser active" signals (or vice versa). If the
"laser active" signals are deactivated (e.g. via
PulseLength = 0), then no switching will occur and
the standby signals will be permanently outputted
even during execution of such commands.
If the standby signals are deactivated, then for "laser
standby" operation all outputs will be set to their
basic TTL states (e.g. LOW if a high signal level was
specified via set_laser_control).
If activated standby signals shall be deactivated when
a list is halted via
pause_list
matically switches off only the laser control signals
for "laser active" operation), then this must be
explicitly induced by calling the command
set_standby(PulseLength = 0). If you do not know
the currently set standby parameters at the moment
you call the control command
(the standby parameters could have been changed
within a list) these can be read out via the control
command get_standby.
or
®
(here the RTC
5 auto-
pause_list
command
129