Section 6: SCPI command reference

*RCL

This command returns the instrument to the setup that was saved with the *SAV command.
Type
Command only
Usage
*RCL <n>
<n>
Details
Restores the state of the instrument from a copy of user-saved settings that are stored in the setup
memory. The settings are saved using the *SAV command.
If you view the user-saved settings from the front panel of the instrument, these are stored as scripts
named Setup0<n>.
Example
*RCL 3
Also see
Saving setups

*SAV

(on page 6-14)
*SAV
This command saves the present instrument settings as a user-saved setup.
Type
Command only
Usage
*SAV <n>
<n>
Details
Save the present instrument settings as a user-saved setup. You can restore the settings with the
*RCL command.
Any command that is affected by *RST can be saved with the *SAV command.
You can save up to 5 user-saved setups. Any settings that had been stored previously as <n> are
overwritten.
If you view the user-saved setups from the front panel of the instrument, they are stored as scripts
named Setup0<n>.
6-14
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
An integer from 0 to 4 that represents the saved setup
(on page 2-150)
Affected by
Not applicable
An integer from 0 to 4
Where saved
Not applicable
Restores the settings stored in memory location 3.
Where saved
Nonvolatile memory
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable
Default value
Not applicable