Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Measure with digitize voltage using the front panel
To make a digitize voltage measurement using the front panel:
1. Make the connections as shown in
2. Press the FUNCTION key.
3. Select the Digitize Functions tab.
4. Select Digitize Voltage.
5. Press the MENU key.
6. Under Measure, select Settings.
7. Select the settings for your application. For descriptions of the options, refer to
measure settings
8. Press the TRIGGER key for 2 seconds and verify that the instrument is set to Continuous
Measurement.
The measurements display on the front panel.
Settings available for digitize voltage measurements
See
Digitize voltage measure settings
digitizing voltage measurements.
Digitize current measurements
The digitize current function makes accurate, predictably spaced current measurements.
Do not apply more than 250 V
caution may result in instrument damage.
Digitize current measure connections
DMM7510-901-01 Rev. B / May 2015
Digitize voltage measure connections
(on page 2-33).
(on page 2-33) for settings that are available when you are
between INPUT LO and the AMPS input. Failure to observe this
peak
Figure 94: Front panel connections: Digitize current measurement
Section 2: General operation
(on page 2-128).
Digitize Voltage
2-129