:CALCulate2:<function>:LIMit<Y>:FAIL?
This command queries the results of a limit test.
Type
Query only
Usage
:CALCulate2:<function>:LIMit<Y>:FAIL?
<function>
<Y>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This command queries the result of a limit test for the selected measurement function.
The response message indicates if the limit test passed or how it failed (on the high or low limit).
If autoclear is set to off, reading the results of a limit test does not clear the fail indication of the test.
To clear a failure, send the clear command. To automatically clear the results, set auto clear on.
If auto clear is set to on and you are making a series of measurements, the last measurement limit
determines the fail indication for the limit. If auto clear is turned off, the results return a test fail if any
of one of the readings failed.
To use this attribute, you must set the limit state to on.
The results of the limit test for limit Y:
NONE: Test passed; the measurement is between the upper and lower limits
HIGH: Test failed; the measurement exceeded the upper limit
LOW: Test failed; the measurement exceeded the lower limit
BOTH: Test failed; the measurement exceeded both limits
6-26
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
The function to which the setting applies; see Functions
Limit number: 1 or 2
RESistance
FRESistance
DIODe
CAPacitance
Where saved
Not applicable
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent

Advertisement

Table of Contents
loading

Table of Contents