Calculate2:<Function>:Limit<Y>:Clear[:Immediate] - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
Example
:CALC2:VOLT:LIM1:CLE:AUTO ON
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
Also see
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
This command clears the results of the limit test defined by Y.
Type
Command only
Usage
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
<function>
<Y>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
Use this command to clear the test results of limit Y when the limit auto clear option is turned off. Both
the high and low test results are cleared.
To avoid the need to manually clear the test results for a limit, turn the auto clear option on.
6-24
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Set limit autoclear on.
Enable the beeper for limit 1 when a voltage measurement exceeds
the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results display on the front
panel
Return the test results; example if the test fails on the low limit:
LOW
The test results are automatically cleared.
Affected by
Not applicable
The function to which the setting applies; see Functions
Limit number: 1 or 2
RESistance
FRESistance
DIODe
CAPacitance
(on page 6-24)
Where saved
Not applicable
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent