Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Details
You cannot assign user-defined reading buffers the name defbuffer1 or defbuffer2.
If you create a reading buffer that has the same name as an existing user-defined buffer, the existing
buffer is overwritten by the new buffer. Any data in the existing buffer is lost.
When you create a reading buffer, it becomes the active buffer. If you create two reading buffers, the
last one you create becomes the active buffer.
The default fill mode of a user-defined buffer is once. You can change it to continuous.
Once the buffer style is selected, it cannot be changed.
Once you store the first reading in a compact buffer, you cannot change certain measurement
settings, including range, display digits, and units; you must clear the buffer first.
Not all remote commands are compatible with the compact, writable, and full writable buffer styles.
Check the Details section of the command descriptions before using them with any of these buffer
styles.
Writable readings are used to bring external data into the instrument. You cannot assign them to
collect data from the instrument.
You can change the buffer capacity for an existing buffer through the front panel or by using the
bufferVar.capacity command.
Example
capTest2 = buffer.make(200, buffer.STYLE_COMPACT)
Creates a 200-element compact reading buffer named capTest2.
Also see
bufferVar.capacity
bufferVar.fillmode
buffer.write.format()
buffer.write.reading()
Reading buffers
Remote buffer operation
Writable reading buffers
DMM7510-901-01 Rev. B / May 2015
(on page 8-22)
(on page 8-28)
(on page 8-43)
(on page 8-45)
(on page 3-13)
(on page 3-30)
(on page 3-34)
Section 8: TSP command reference
8-19