Dmm.digitize.limit[Y].Fail - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

dmm.digitize.limit[Y].fail

This attribute queries the results of a limit test.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
value = dmm.digitize.limit[Y].fail
value
Y
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This command queries the result of a limit test for the selected digitize function.
The response message indicates if the limit test passed or how it failed (on the high or low limit).
If autoclear is set to off, reading the results of a limit test does not clear the fail indication of the test.
To clear a failure, send the clear command. To automatically clear the results, set auto clear on.
If auto clear is set to on and you are making a series of measurements, the last measurement limit
determines the fail indication for the limit. If auto clear is turned off, the results return a test fail if any
of one of the readings failed.
To use this attribute, you must set the limit state to on.
If the readings are stored in a reading buffer, you can use the bufferVar.statuses command to
see the results.
Example
This example enables limits 1 and 2 for digitize voltage measurements. Limit 1 is checking for
readings to be between 3 and 5 V, while limit 2 is checking for the readings to be between 1 and 7 V.
The auto clear feature is disabled, so if any reading is outside these limits, the corresponding fail is 1.
Therefore, if any one of the fails is 1, analyze the data in the reading buffer to determine which
reading failed the limits.
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Not applicable
Not applicable
The results of the limit test for limit Y:
dmm.FAIL_NONE: Test passed; measurement under or equal to the high limit
dmm.FAIL_HIGH: Test failed; measurement exceeded high limit
dmm.FAIL_LOW: Test failed; measurement exceeded low limit
dmm.FAIL_BOTH: Test failed; measurement exceeded both limits
Limit number: 1 or 2
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
Section 8: TSP command reference
Default value
Not applicable
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
8-97