Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
[:SENSe[1]]:<function>:COUPling:AC:FREQuency
This command allows you to optimize the amplitude to compensate for signal loss across the coupling capacitor
when AC coupling is selected.
Type
Command and query
Usage
[:SENSe[1]]:<function>:AC:FREQuency <range>
[:SENSe[1]]:<function>:AC:FREQuency?
<function>
<range>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This command is only used when the digitize coupling type is set to AC.
For example, if you are measuring a 50 Hz signal, you could set this to 50 Hz to compensate for
voltage drop across the coupling capacitor.
Example
DIG:VOLT:COUP AC
DIG:VOLT:COUP:AC:FILT FAST
DIG:VOLT:COUP:AC:FREQ 500
Also see
[:SENSe[1]]:<function>:COUPling
DMM7510-901-01 Rev. B / May 2015
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The frequency: 3 Hz to 1 MHz
RESistance
FRESistance
DIODe
CAPacitance
(on page 6-75)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the coupling type to AC.
Set the filter to fast.
Set the frequency to 500 Hz.
Section 6: SCPI command reference
Default value
1000
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-77