[:Sense[1]]:<Function>:Bias:actual - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
[:SENSe[1]]:<function>:BIAS:ACTual?
This command returns the amount of current the instrument is sourcing when it makes measurements.
Type
Query only
Usage
[:SENSe[1]]:<function>:BIAS:ACTual?
<function>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
Returns the bias level: 10 µA, 100 µA, or 1 mA.
Reads the actual amount of current that is sourced by the instrument when a measurement is made.
Example
:FUNC "DIOD"
:DIOD:BIAS:ACTUAL?
Also see
[:SENSe[1]]:<function>:BIAS:LEVel
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
The function to which the setting applies; see Functions
RESistance
FRESistance
DIODe
CAPacitance
(on page 6-74)
Where saved
Not applicable
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the function to diode.
Query the actual bias level that is used; example
output:
0.00100351
Section 6: SCPI command reference
Default value
Not applicable
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-73

Advertisement

Table of Contents
loading

Table of Contents