Acal.lastrun.tempdiff - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

acal.lastrun.tempdiff

This attribute returns the difference between the internal temperature and the temperature when auto calibration
was last run.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
temperature = acal.lastrun.tempdiff
temperature
Details
The temperature is displayed in Celsius (°C).
The instrument updates the internal temperature value when the instrument refreshes autozero. If
autozero is set to off or if autozero is not available for the selected function (such as capacitance,
continuity, frequency or period), the internal temperature value is not updated.
Example
print(acal.lastrun.tempdiff)
Also see
acal.lastrun.internaltemp
acal.run()
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
The internal temperature
(on page 8-8)
(on page 8-12)
Where saved
Not applicable
Returns the difference between the temperature of the
instrument when auto calibration was run and the present
internal temperature.
Example output:
4.5678
Section 8: TSP command reference
Default value
Not applicable
8-9