[:Sense[1]]:<Function>:Average:tcontrol - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
Also see
Filtering measurement data
[:SENSe[1]]:<function>:AVERage:COUNt
[:SENSe[1]]:<function>:AVERage:TCONtrol
[:SENSe[1]]:<function>:AVERage:WINDow
[:SENSe[1]]:<function>:AVERage:TCONtrol
This command sets the type of averaging filter that is used for the selected measure function when the
measurement filter is enabled.
Type
Command and query
Usage
[:SENSe[1]]:<function>:AVERage:TCONtrol <type>
[:SENSe[1]]:<function>:AVERage:TCONtrol?
<function>
<type>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
DMM7510-901-01 Rev. B / May 2015
(on page 3-11)
(on page 6-67)
(on page 6-69)
(on page 6-71)
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The filter type to use when filtering is enabled; set to one of the following values:
Repeating filter: REPeat
Moving filter: MOVing
RESistance
FRESistance
DIODe
CAPacitance
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Section 6: SCPI command reference
Default value
REP
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-69