Section 6: SCPI command reference

:ACAL:RUN

This command immediately runs auto calibration and stores the constants.
Type
Command only
Usage
:ACAL:RUN
Details
During auto calibration, a progress message is displayed on the front panel. At completion, an event
message is generated.
If you have set up auto calibration to run at a scheduled interval, when you send the run command,
the instrument adjusts the next scheduled auto calibration to be the next interval. For example, if auto
calibration is scheduled to run every 7 days, but you run auto calibration on day 3, the next auto
calibration will run 7 days after day 3.
Example
ACAL:RUN
Also see
:ACAL:SCHedule
Auto calibration
6-20
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
(on page 6-21)
(on page 3-44)
Where saved
Nonvolatile memory
Auto calibration starts running.
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable