To change the IP configuration using the instrument web page:
1. Access the internal web page as described in
interface
2. From the navigation bar on the left, in the LXI Home menu, select IP Config.
Click Modify. The Modify IP Configuration page is displayed.
3. Change the values.
4. Click Submit. The instrument reconfigures its settings, which may take a few moments.
You may lose your connection with the web interface after clicking Submit. This is normal and does
not indicate an error or failure of the operation. If this occurs, find the correct IP address and reopen
the web page of the instrument to continue.
Review events in the event log
The event log records all LXI events that the instrument generates and receives. The log includes the
following information:
The EventID column, which shows the event identifier that generated the event.
The System Timestamp column, which displays the seconds and nanoseconds when the event
occurred.
The Data column, which displays the text of the event message.
To clear the event log and update the information on the screen, click the Refresh button.
DMM7510-901-01 Rev. B / May 2015
(on page 2-82).
Figure 53: Modify IP Configuration web page