Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

dmm.measure.analogtrigger.mode

This attribute configures the type of signal behavior that can generate an analog trigger event.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
setting = dmm.measure.analogtrigger.mode
dmm.measure.analogtrigger.mode = setting
setting
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
When edge is selected, the analog trigger occurs when the signal crosses a certain level. You also
specify if the analog trigger occurs on the rising or falling edge of the signal.
When pulse is selected, the analog trigger occurs when a pulse passes through the specified level
and meets the constraint that you set on its width. You also specify the polarity of the signal (above or
below the trigger level).
When window is selected, the analog trigger occurs when the signal enters or exits the window
defined by the low and high signal levels.
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
The mode setting:
Edge (signal crosses one level): dmm.MODE_EDGE
Pulse (two complementary edge events meet a specified time constraint):
dmm.MODE_PULSE
Window (signal enters or exits a window defined by two levels):
dmm.MODE_WINDOW
No analog triggering: dmm.MODE_OFF
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
Section 8: TSP command reference
Default value
dmm.MODE_OFF
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
8-121