:TRACe:SAVE

This command saves data from the specified reading buffer to a USB flash drive.
Type
Command only
Usage
:TRACe:SAVE "<fileName>"
:TRACe:SAVE "<fileName>", "<bufferName>"
:TRACe:SAVE "<fileName>", "<bufferName>", <timeFormat>
:TRACe:SAVE "<fileName>", "<bufferName>", <timeFormat>, <start>, <end>
<fileName>
<bufferName>
<timeFormat>
<start>
<end>
Details
The filename must specify the full path (including /usb1/). If included, the file extension must be set
to .csv (if no file extension is specified, .csv is added).
For options that save more than one item of time information, each item is comma-delimited. For
example, the default format is date, time, and fractional seconds for each reading.
The Model DMM7510 does not check for existing files when you save. Verify that you are using a
unique name to avoid overwriting any existing .csv files on the flash drive.
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
A string that indicates the name of the file on the USB flash drive in which to save
the reading buffer
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Defines how date and time information from the buffer is saved in the file on the
USB flash drive; the values are:
Dates, times, and fractional seconds are saved; the default value: FORMat
Relative timestamps are saved: RELative
Seconds and fractional seconds are saved: RAW
Timestamps are saved: STAMp
Defines the starting point in the buffer to start saving data
Defines the ending point in the buffer to stop saving data