Section 8: TSP command reference
Example
reset()
trigger.model.load("SimpleLoop", 10, 0, defbuffer1)
display.prompt(display.BUTTONS_YESNO, "Would you like to make 10 DC voltage
readings now?")
promptID, result = display.waitevent()
if result == display.BUTTON_YES then
trigger.model.initiate()
end
display.prompt(display.BUTTONS_YESNO, "Would you like to switch to the Graph
screen?")
promptID, result = display.waitevent()
if result == display.BUTTON_YES then
display.changescreen(display.SCREEN_GRAPH)
end
Create a simple loop that will make 10 measurements and save them in default buffer 1.
Display the prompt shown here:
If the user presses Yes, the measurements are made.
If the user presses No, the measurements are not made and the message is removed.
Display the prompt "Would you like to switch to the Graph screen?"
If the user presses Yes, the Graph screen is displayed.
If the user presses No, the user remains on the present screen.
Also see
display.delete()
display.waitevent()

display.readingformat

This attribute determines the format that is used to display measurement readings on the front-panel display of
the instrument.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
format = display.readingformat
display.readingformat = format
format
8-68
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 8-58)
(on page 8-70)
Affected by
Not applicable
Use exponent format: display.FORMAT_EXPONENT
Add a prefix to the units symbol, such as k, m, or µ: display.FORMAT_PREFIX
Where saved
Default value
Nonvolatile memory
display.FORMAT_PREFIX