Trigger.timer[N].Delay; Trigger.timer[N].Delaylist - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference

trigger.timer[N].delay

This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
interval
N
Details
Once the timer is enabled, each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[N].delaylist = {interval}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
If you use the trigger timer with a trigger model, make sure the trigger timer delay is set so that the
readings are paced correctly.
Example
trigger.timer[1].delay = 50e-6
Also see
trigger.timer[N].reset()

trigger.timer[N].delaylist

This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
intervals = trigger.timer[N].delaylist
trigger.timer[N].delaylist = intervals
intervals
N
8-330
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger timer N reset
Delay interval in seconds (8 µs to 100 ks)
Trigger timer number (1 to 4)
(on page 8-333)
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger timer N reset
Table of delay intervals in seconds
Trigger timer number (1 to 4)
Where saved
Default value
10e-6 (10 µs)
Configuration script
Set the trigger timer 1 to delay for
50 µs.
Where saved
Default value
Configuration script
{10e-6}
DMM7510-901-01 Rev. B / May 2015