Keithley DMM7510 Reference Manual page 690

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
dmm.digitize.analogtrigger.pulse.level
This attribute defines the pulse level that generates an analog trigger event.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = dmm.digitize.analogtrigger.pulse.level
dmm.digitize.analogtrigger.pulse.level = value
value
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
Only available when the analog trigger mode is set to pulse.
Example
dmm.digitize.func = dmm.FUNC_DIGITIZE_VOLTAGE
dmm.digitize.analogtrigger.mode = dmm.MODE_PULSE
dmm.digitize.analogtrigger.pulse.level = 5
dmm.digitize.analogtrigger.pulse.width = 30e-6
dmm.digitize.analogtrigger.pulse.condition = dmm.CONDITION_LESS
dmm.digitize.analogtrigger.pulse.polarity = dmm.POLARITY_BELOW
Set function to digitize voltage.
Set the analog trigger mode to pulse.
Set the analog trigger level to 5 V.
Set the analog trigger pulse width to 30 µs.
Set the condition to be detect trigger within the pulse width.
Set the trigger to occur when the pulse is below the level.
Also see
Analog triggering overview
dmm.digitize.analogtrigger.mode
dmm.digitize.func
dmm.digitize.analogtrigger.pulse.condition
dmm.digitize.analogtrigger.pulse.polarity
dmm.digitize.analogtrigger.pulse.width
dmm.measure.analogtrigger.pulse.level
DMM7510-901-01 Rev. B / May 2015
Affected by
Restore configuration
Instrument reset
Power cycle
Measure configuration list
The signal level
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 3-64)
(on page 8-74)
(on page 8-90)
(on page 8-76)
(on page 8-78)
(on page 8-79)
(on page 8-124)
Section 8: TSP command reference
Where saved
Default value
Configuration script
0
Measure configuration list
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
8-77