Trigger:blender<N>:Stimulus<M - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
:TRIGger:BLENder<n>:STIMulus<m>
This command specifies the events that trigger the blender.
Type
Command and query
Usage
:TRIGger:BLENder<n>:STIMulus<m> <event>
:TRIGger:BLENder<n>:STIMulus<m>?
<n>
<m>
<event>
Details
There are four stimulus inputs that can each select a different event.
Use zero to disable the blender input.
The <event> parameter may be any of the trigger events shown in the following table.
Trigger events
Event description
No trigger event
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line <n> (1 to 6)
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object
<n> (1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events BLENder<n>
Trigger timer <n> (1 to 4) expired
Analog trigger
External in trigger
6-180
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Trigger blender clear
The blender number (1 or 2)
The stimulus input number (1 to 4)
See Details
Where saved
Save settings
Event constant
NONE
DISPlay
NOTify<n>
COMMand
DIGio<n>
TSPLink<n>
LAN<n>
TIMer<n>
ATRigger
EXTernal
DMM7510-901-01 Rev. B / May 2015
Default value
NONE