Trigger.blender[N].Clear(); Trigger.blender[N].Orenable - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference

trigger.blender[N].clear()

This function clears the blender event detector and resets the overrun indicator of blender N.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.blender[N].clear()
N
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Also see
None

trigger.blender[N].orenable

This attribute selects whether the blender performs OR operations or AND operations.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
N
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
events (AND) before signaling an output event.
8-254
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Where saved
The blender number (1 or 2)
Clears the event detector for blender 2.
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Power cycle
Trigger blender N reset
The type of operation:
true: OR operation
false: AND operation
The blender number (1 or 2)
Default value
Default value
false (AND)
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading

Table of Contents