Capacitance measurements

With a Model DMM7510, you can measure capacitance.
The capacitance function sources a constant I
current through the device under test (DUT) while
test
measuring voltage (dV) in a fixed time internal (dt). The Capacitance measurement is:
I
* dt / dV
test
Capacitance measurements have two measurement phases, discharge and charge. During the
discharge phase, the DUT is connected through an internal 13 mA current source and discharged to
approximately 0 V. In the charge phase, the I
is sourced while measuring the voltage. If the voltage
test
on the DUT exceeds 2.8 V ±10 %, the I
is halted and the voltage is held until the discharge phase.
test
If the voltage is less than 2.8 V, the resultant capacitance measurement is calculated.
Capacitance supports 1 nF to 1 mF ranges. Each range measures from 0 % to 120 % full scale.
Reading rates vary based on range and the percent of full scale.
The 13 mA discharge and I
currents are protected to 1000 V.
test
Capacitance has a fixed aperture time.
Do not apply more than 1000 V
between INPUT HI and LO. Failure to observe this caution may
peak
result in instrument damage.
Capacitance measure connections
Figure 86: Front panel connections: Capacitor measurement
2-122
DMM7510-901-01 Rev. B / May 2015