Section 6: SCPI command reference
Inside: The measurement is inside the values set by limits A and B; limit A must be the low value
and Limit B must be the high value
Outside: The measurement is outside the values set by limits A and B; limit A must be the low
value and Limit B must be the high value
The measurement block must be a measure or digitize building block that occurs in the trigger model
before the branch-on-constant-limits block. The last measurement from a measure or digitize building
block is used.
If the limit A is more than the limit B, the values are automatically swapped so that the lesser value is
used as the lower limit.
Example
TRIGger:BLOCk:BRANch:LIMit:CONStant 5, OUTside, 0.15, 0.65, 8
Configure trigger block 5 to check for measurements in the last measure or digitize block. If the
measurements are outside of the 0.15 and 0.65 limits, branch to block 8.
Also see
Constant Limit block

:TRIGger:BLOCk:BRANch:LIMit:DYNamic

This command defines a trigger model block that goes to a specified block in the trigger model if a measurement
meets user-defined criteria.
Type
Command only
Usage
:TRIGger:BLOCk:BRANch:LIMit:DYNamic <blockNumber>, <limitType>, <limitNumber>,
<branchToBlock>
:TRIGger:BLOCk:BRANch:LIMit:DYNamic <blockNumber>, <limitType>, <limitNumber>,
<branchToBlock>, <measureBlock>
<blockNumber>
<limitType>
<limitNumber>
<branchToBlock>
<measureBlock>
6-188
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 3-85)
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The type of limit (ABOVe, BELow, INside, or OUTside)
The limit number (1 or 2)
The block number of the trigger model block to execute when the limits
are met
The block number of the measure or digitize block that makes the
measurements to be compared; if this is 0 or undefined, the trigger model
uses the previous measure or digitize block
Where saved
Save settings
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable