Dmm.measure.fourrtd - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference

dmm.measure.fourrtd

This attribute defines the type of 4-wire RTD that is being used
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
RTDType = dmm.measure.fourrtd
dmm.measure.fourrtd = RTDType
RTDType
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
The transducer type must be set to temperature and the transducer must be set to 4-wire RTD before
you can set the RTD type.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_FOURRTD
dmm.measure.fourrtd =
Also see
dmm.measure.rtdalpha
dmm.measure.rtdbeta
dmm.measure.rtddelta
dmm.measure.rtdzero
dmm.measure.transducer
Temperature measurements
8-154
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
Measure configuration list
The type of 4-wire RTD:
PT100: dmm.RTD_PT100
PT385: dmm.RTD_PT385
PT3916: dmm.RTD_PT3916
D100: dmm.RTD_D100
F100: dmm.RTD_F100
User-specified type: dmm.RTD_USER
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
dmm.RTD_PT3916
(on page 8-184)
(on page 8-185)
(on page 8-186)
(on page 8-187)
(on page 8-197)
(on page 2-120)
Where saved
Default value
Configuration script
dmm.RTD_PT100
Measure configuration list
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Set the measure function to temperature.
Set the transducer type to 4-wire RTD.
Set the RTD type to PT3916.
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading