Program Fragment
PRINT #1, "output 16; :syst:pres"
PRINT #1, "output 16; :init:cont off;:abort"
PRINT #1, "output 16; :trig:coun 1; sour tim"
PRINT #1, "output 16; :samp:coun 5"
PRINT #1, "output 16; :init; *opc?"
PRINT #1, "enter 16"
LINE INPUT #2, a$
PRINT a$
*OPT? — Option Identification Query
Description
The response message indicates the presence or absence of an optional scanner card. For
example:
0
200X-SCAN
*RCL — Recall
Parameters
<NRf>=0
Description
Use this command to return the Model 2010 to the configuration stored in memory. The *SAV
command is used to store the setup configuration in memory location.
Only one setup configuration can be saved and recalled.
The Model 2010 ships from the factory with :SYSTen:PRESet defaults loaded into the
available setup memory. If a recall error occurs, the setup memory defaults to the
:SYSTem:PRESet values.
No scanner card installed
Scanner card installed
Remote Operation
'Select defaults
'Place 2010 in idle
'Program for five measurements
'and stop (idle)
'Start measurements and send
'*opc?
'Get response when 2010 goes into
'idle
'Read contents of Output Queue
'Display the ASCII "1"
Determine if an option is installed
Return to setup stored in memory
4-41