[:Sense[1]]:<Function>:Line:sync - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
[:SENSe[1]]:<function>:LINE:SYNC
This command determines if line synchronization is used during the measurement.
Type
Command and query
Usage
[:SENSe[1]]:<function>:LINE:SYNC <state>
[:SENSe[1]]:<function>:LINE:SYNC?
<function>
<state>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
When line synchronization is enabled, measurements are initiated at the first positive-going zero
crossing of the power line cycle after the trigger.
Example
CURR:LINE:SYNC ON
Also see
Line cycle synchronization
6-84
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
Disable: OFF or 0
Enable: ON or 1
RESistance
FRESistance
DIODe
CAPacitance
(on page 4-1)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Turn on line synchronization when DC current is
measured.
DMM7510-901-01 Rev. B / May 2015
Default value
0 (OFF)
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent