[:SENSe[1]]:<function>:ATRigger:MODE
This command configures the type of signal behavior that can generate an analog trigger event.
Type
Command and query
Usage
[:SENSe[1]]:<function>:ATRigger:MODE <setting>
[:SENSe[1]]:<function>:ATRigger:MODE?
<function>
<setting>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
When edge is selected, the analog trigger occurs when the signal crosses a certain level. You also
specify if the analog trigger occurs on the rising or falling edge of the signal.
When pulse is selected, the analog trigger occurs when a pulse passes through the specified level
and meets the constraint that you set on its width. You also specify the polarity of the signal (above or
below the trigger level).
When window is selected, the analog trigger occurs when the signal enters or exits the window
defined by the low and high signal levels.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE EDGE
CURR:ATR:EDGE:LEV 2.5
CURR:ATR:EDGE:SLOP RIS
Also see
Analog triggering overview
DMM7510-901-01 Rev. B / May 2015
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The setting:
Edge (signal crosses one level): EDGE
Pulse (two complementary edge events meet a specified time constraint):
PULse
Window (signal enters or exits a window defined by two levels): WINDow
No analog triggering: OFF
RESistance
FRESistance
DIODe
CAPacitance
(on page 3-64)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
Set the analog trigger slope to rising.
Section 6: SCPI command reference
Default value
OFF
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-59