Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

Establishing a relative offset value

You can use the Model DMM7510 to automatically determine the relative offset, or you can assign a
specific relative offset value.
Automatically acquiring a relative offset value
When you automatically acquire a relative offset value, the Model DMM7510:
Makes a new measurement.
Stores the measurement as the new relative offset level.
Before acquiring the offset, apply the signal that you want to offset the measurement by.
Using the front panel:
1. Press the FUNCTION key and select the measure function.
2. Press the MENU key.
3. Select Calculations.
4. For Rel, select Acquire. The relative offset value is displayed to the right.
Selecting Acquire from the front panel automatically enables the relative offset value, unless an
overflow reading is detected.
You can also enable or disable the relative offset feature through the SETTINGS swipe screen Rel
option.
Using a remote interface:
SCPI commands: Refer to
[:SENSe[1]]:<function>:RELative:STATe
TSP commands: Refer to
dmm.measure.rel.enable
When the relative offset is selected, the REL annunciator to the right of the measurement is
displayed.
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:RELative:ACQuire
(on page 6-96).
dmm.measure.rel.acquire()
(on page 8-180).
Section 3: Functions and features
(on page 6-94) and
(on page 8-179) and