Section 6: SCPI command reference

:ACAL:LASTrun:TIME?

This command returns the date and time when auto calibration was last run.
Type
Query only
Usage
:ACAL:LASTrun:TIME?
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
MM/DD/YYYY is the month, date, and year
HH:MM:SS.NNNNNNNNN is the hour, minute, second, and fractional second
Example
ACAL:LAST:TIME?
Also see
:ACAL:RUN
(on page 6-20)
6-18
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
Where saved
Nonvolatile memory
Returns the date and time when auto calibration
was last run.
Example output:
08/11/2014 16:30:26.745369595