Keithley DMM7510 Reference Manual page 626

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
acal.schedule()
This function sets how often auto calibration occurs or prompts you to run it.
Type
TSP-Link accessible
Function
Yes
Usage
action, interval, hour = acal.schedule()
acal.schedule(acal.ACTION_NONE)
acal.schedule(action, interval)
acal.schedule(action, interval, hour)
action
interval
hour
Details
Autocalibration does not start until all actions that are active on the instrument are complete. When
the scheduled time occurs, the autocalibration run command is placed in the command queue and
will be executed after any previously sent commands or actions have executed. For example, if a
trigger model is running when autocalibration is scheduled to run, autocalibration does not start until
the trigger model stops.
If there is a command or action that is waiting a long time for an event, the autocalibration will not run
until the event occurs, the action is aborted, or the instrument power is cycled.
If the scheduled time for autocalibration occurs before the warmup period completes, the instrument
will not start autocalibration. The instrument waits until the warmup period is complete before starting
a scheduled autocalibration. A message is displayed when warmup is complete and autocalibration is
going to run.
If the instrument is powered off when an autocalibration was scheduled, autocalibration is run as soon
as the warmup period is complete when the instrument is powered on.
You can run autocalibration manually even if a scheduled autocalibration is set.
When autocalibration is scheduled to run at a scheduled interval, but it runs at a time other than the
scheduled interval, subsequent scheduled intervals are adjusted according to the actual
autocalibration start time.
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Restore configuration
Nonvolatile memory
Configuration script
Determines when and if the instrument automatically runs auto calibration:
To run auto calibration at the scheduled time: acal.ACTION_RUN
To notify you that the auto calibration needs to be run at the scheduled time:
acal.ACTION_NOTIFY
To turn off scheduling: acal.ACTION_NONE; no other parameters are
needed if none is selected
Determines how often auto calibration should be run or notification should occur:
Every 8 hours: acal.INTERVAL_8HR
Every 16 hours: acal.INTERVAL_16HR
Every day: acal.INTERVAL_1DAY
Every 7 days: acal.INTERVAL_7DAY
Every 14 days: acal.INTERVAL_14DAY
Every 30 days: acal.INTERVAL_30DAY
Every 90 days: acal.INTERVAL_90DAY
Specify when the auto calibration should occur; specify in 24-hour time format (0 to
23; default is 0); not available for the 8-hour or 16-hour interval
Section 8: TSP command reference
Default value
Run every 8 hours starting at
midnight
8-13