[:SENSe[1]]:<function>:RANGe[:UPPer]
This command determines the positive full-scale measure range.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RANGe[:UPPer] <n>
[:SENSe[1]]:<function>:RANGe[:UPPer] DEFault
[:SENSe[1]]:<function>:RANGe[:UPPer] MINimum
[:SENSe[1]]:<function>:RANGe[:UPPer] MAXimum
[:SENSe[1]]:<function>:RANGe[:UPPer]?
[:SENSe[1]]:<function>:RANGe[:UPPer]? DEFault
[:SENSe[1]]:<function>:RANGe[:UPPer]? MINimum
[:SENSe[1]]:<function>:RANGe[:UPPer]? MAXimum
<function>
<n>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
You can assign any real number using this command. The instrument selects the closest fixed range
that is large enough to measure the entered number. For example, for current measurements, if you
expect a reading of approximately 9 mA, set the range to 9 mA to select the 10 mA range. When you
read this setting, you see the positive full-scale value of the measurement range that the instrument is
presently using.
This command is primarily intended to eliminate the time that is required by the instrument to
automatically search for a range.
When a range is fixed, any signal greater than the entered range generates an overrange condition.
When an overrange condition occurs, the front panel displays "Overflow" and the remote interface
returns 9.9e+37.
When you set a value for the measurement range, the measurement autorange setting is
automatically disabled for the selected measurement function.
The range for measure functions defaults to autorange. When you switch from autorange to range,
the range is set to the last selected autorange value.
6-90
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
Set this command to a specific value or a preset value:
See Details
RESistance
FRESistance
DIODe
CAPacitance
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent

Advertisement