Format:border - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

:FORMat:BORDer

This command sets the byte order for the IEEE-754 binary formats.
Type
Command and query
Usage
:FORMat:BORDer <name>
:FORMat:BORDer?
<name>
Details
This attribute selected the byte order in which data is written.
The SWAPped byte order must be used when transmitting binary data to a computer with a Microsoft
Windows operating system.
The ASCII data format can only be sent in the normal byte order. If the ASCII format is selected, the
SWAPped selection is ignored.
When you select NORMal byte order, the data format for each element is sent as follows:
Byte 1 Byte 2 Byte 3 Byte 4
(Single precision)
When you select SWAPped, the data format for each element is sent as follows:
Byte 4 Byte 3 Byte 2 Byte 1
(Single precision)
The #0 header is not affected by this command. The header is always sent at the beginning of the
data string for each measurement conversion.
Example
FORM:BORD NORM
Also see
:FORMat[:DATA]
DMM7510-901-01 Rev. B / May 2015
Affected by
Recall settings
Instrument reset
Power cycle
The binary byte order:
Normal byte order: NORMal
Reverse byte order for binary formats: SWAPped
(on page 6-50)
Section 6: SCPI command reference
Where saved
Save settings
Use the normal byte order.
Default value
SWAP